Method

Parameters

The parameters for the command

MEAS pH

are configured on the following 4 tabs:

General/Hardware

Parameters for device, sensor and stirrer.

Measuring parameters

Parameters for the measurement.

Evaluations

Defining further methods for evaluating measurement curves.

Additional measured values

Defining further measured values which can be saved as additional col-
umns in the list of measuring points.

General/Hardware

The general parameters for the control device, the sensor and the stirrer are de-
fined on this tab.

Command name

25 characters, [ MEAS pH # ]

Name of the command.

Device

Device name

Device name, [ not defined ]

Selection of a device from the device table. Only those devices being able
to execute the command are selectable. Additionally,

not defined

can be se-

lected. In this case, the device name must be assigned by the user at the
start of the method.

Device type

Device types, [ Titrando ]

Display or selection of the device type. If a device is selected under

Device

name

, this parameter is not editable any more. The device type belonging to

the selected device is displayed instead.
If the option

not defined

is selected as

Device name

, all device types being

able to execute the command can be selected independently from the de-
vices in the device table.