352

Method

Lower limit

[ -2000.0 ] ... 2000.0 mV

(for

Window

=

Measured value

)

[ 0.00000 ] ... 9999.99 mL

(for

Window

=

Volume

)

[ 0 ] ... 999999 s

(for

Window

=

Time

)

Value for the lower limit of the window.

Upper limit

-2000.0 ... [ 2000.0 ] mV

(for

Window

=

Measured value

)

0.00000 ... [ 9999.99 ] mL

(for

Window

=

Volume

)

0 ... [ 999999 ] s

(for

Window

=

Time)

Value for the upper limit of the window.

Gran evaluation

on, [ off ]

(only Titrando)

With the Gran evaluation titration curves are evaluated according to the

Gran plot procedure (see Titration commands - Evaluation - Gran evalua-
tion
).

Initial volume

0.01 ... [ 50.00 ] ... 9999.99 mL

Volume which must be present in the measuring vessel before the com-
mand is started.

Lower limit

[ -2000.0 ] ... 2000.0 mV

Value for the lower limit of the Gran evaluation range.

Upper limit

-2000.0 ... [ 2000.0 ] mV

Value for the upper limit of the Gran evaluation range.

Additional measured values

Screenshot and parameters: see DET pH - Additional measured values

DET Ipol

Command for Dynamic Equivalence Point Titrations with voltametric meas-
urement with (selectable polarization current).

Devices

This command can be executed with the following devices which have the DET
mode:

Titrino: 716, 721, 736, 751, 785, 794, 798, 799

Titrando: 808, 809, 835, 836, 855, 857

Appearance

The command has the following appearance:

Parameters

The parameters for the command

DET Ipol

are configured in the following 7 tabs:

General/Hardware

Parameters for devices, dosing devices, sensors and stirrers.