Metrohm tiamo 1.2 Manual Benutzerhandbuch

Seite 392

Advertising
background image

366

Method

der

Sensor

is set to

automatic

or

continuous

then the temperature will be

measured continuously.

Stop conditions

Conditions for stopping the titration. The stop conditions are checked at the same
time and the condition which is fulfilled first stops the titration.

Command name

25 characters, [ DET Upol # ]

Name of the command.

Stop volume

0.00000 ... [ 100.000 ] ... 9999.99 mL, off

(Titrando)

0.000 ... [ 100.000 ] ... 9999.99 mL, off

(Titrino)

Stops when the given volume (including start conditions) has been added
after the start of the titration. The stop volume should be adapted to suit the
sample weight or the titration vessel size.

Stop measured value

-200.0 ... 200.0 µA, [ off ]

Stops when the preset value for a measuring point has been exceeded or

undercut since the start of the titration.

Stop EP

1 ... [ 9 ], off

Stops when the given number of EPs has been found..

Volume after EP

0.01000 ... 9999.99 mL, [ off ]

(only Titrando)

When the number of equivalence points defined under

Stop EP

has been

found this volume will be added. In this way the curve can be continued af-
ter the equivalence point has been reached.

Stop time

0 ... 999999 s, [ off ]

(only Titrando)

Stops when the preset time (including start conditions) has elapsed since
the start of the titration.

Filling rate

0.01 ... 166.00 mL/min, [ maximum ]

(Titrando)

0.01 ... 150.00 mL/min, [ maximum ]

(Titrino)

Speed with which the buret is to be refilled after the titration. The maximum
filling rate depends on the cylinder volume of the exchange unit or dosing

unit used.

Advertising