374

Method

Titration parameters

Parameters defining the run of the titration.

Command name

25 characters, [ MET pH # ]

Name of the command.

Titration rate

Titration rate

Three predefined sets of parameters can be selected for the titration rate:

slow

,

optimal

and

fast

. For these sets the parameters for the

Measured value

acceptance

and

Dosing of increments

are not displayed. In order to edit these

parameters the value

user

must be selected.

slow

For titrations in which the finest details are also to be visible. However, this

can also increase noise and lead to the production of unwanted EPs.

[ optimal ]

Parameter set for all standard titrations; optimized for the most frequent
applications.

fast

For less critical rapid titrations.

user

Editing the individual titration parameters which affect the titration rate.

Measured value acceptance

(only visible for

Titration rate

=

user

)

Signal drift

0.1 ... [ 20.0 (slow) ] ... [ 50.0 (optimal) ] ... [ 80.0 (fast) ] ... 999.0 mV/min

(Ti-

trando)

0.5 ... [ 20.0 (slow) ] ... [ 50.0 (optimal) ] ... [ 80.0 (fast) ] ... 999.0 mV/min

(Ti-

trino)

off

Drift for accepting the measured value.

off

means that the measured value

will be accepted when the maximum waiting time has elapsed.