Method

Commands

Depending on the measured value one of the following four

SET

commands can

be chosen:

SET pH

Potentiometric pH measurement with pH electrodes (measured quantity
pH).

SET U

Potentiometric voltage measurement with metal electrodes (measured
quantity voltage U).

SET Ipol

Voltametric measurement with selectable polarization current (measured
quantity voltage U).

SET Upol

Amperometric measurement with selectable polarization voltage (measured
quantity current I).

Control range

Set a large control range for steep curves and a small control range for flat ones.
A good approximation for the start of the control range is given by the point where
the tangents intersect.

SET pH

Command for Set Endpoint Titrations with potentiometric pH measurement.

Devices

This command can be executed with the following devices which have the SET
mode:

Titrino: 702, 716, 718, 719, 720, 721, 736, 751, 758, 785, 794, 798, 799
Titrando: 808, 809, 835, 836, 841, 842, 855, 857

Appearance

The command has the following appearance:

without conditioning with conditioning