Commands

499

Switch off automatically

[ on ], off

The stirrer is automatically switched off at the end of the command if this

option is enabled.

Standard addition

The standard addition parameters are defined on this tab.

Command name

25 characters, [ STDADD dos # ]

Name of the command.

Number of additions

1 ... [ 3 ] ... 19

Number of times the standard solution is to be added.

Volume auxiliary solution

[ 0.000 ] ... 9999.999 mL

Volume of the auxiliary solution (e.g. buffer) to be added to the sample be-
fore the first standard addition.

Addition mode

Fixed volume increments

[ on ], off

If this option is enabled a fixed volume increment muss be added for each
standard addition.

Addition volume

[ 0.00000 ] ... [ 0.100 ] ... 99999.9 mL

Quantity of the fixed volume increment.

Variable volume increments

on, [ off ]

If this option is enabled the defined volume increment muss be added for
each standard addition.

Addition volume 1...19

[ 0.00000 ] ... [ 0.100 ] ... 99999.9 mL

Quantity of the defined volume increments.

Addition solution

(only displayed for

Solution

=

not defined

)

Concentration

0.001 ...[ 1.000 ] ... 999999.999

Concentration of addition solution.