Metrohm tiamo 1.2 Manual Benutzerhandbuch

Seite 543

Advertising
background image

Commands

517

DOS

Command for controlled dosing with possibility for monitoring of measured
value and temperature.

Principle

A predefined volume of a solution is dosed with the DOS command. The meas-
ured value and the temperature can be monitored simultaneously. Three different

types of dosing modes can be chosen by defining two parameters from the fol-
lowing:

Volume

,

Dosing rate

and

Dosing time

. The third parameter is calculated ac-

cording to the following formula:

Volume = Dosing time

Dosing rate

If continuous dosing is required, a tandem dosing setup can be used i. e. dosing
is carried out with a combination of two dosing devices. The second dosing de-
vice is dosing while the first one is being filled and vice versa.

Commands

Depending on the measured value one of the following two

DOS

commands can

be chosen:

DOS pH

Controlled dosing with measurement quantity pH.

DOS U

Controlled dosing with measurement quantity U.

DOS pH

Command for controlled dosing with measured value pH.

Devices

This command can be executed with the following devices which have the DOS
pH
mode:

Titrino: 718, 736, 751, 799
Titrando: 835, 836, 842, 855, 857

Appearance

The command has the following appearance:

Parameters

The parameters for the command

DOS pH

are configured in the following 5 tabs:

General/Hardware

Parameters for devices, dosing devices, sensors and stirrers.

Dosing parameters

Parameters for the dosing settings.

Stop conditions

Definition of conditions which cause the stop of the dosing.

Monitoring

Definition of measured value and temperature monitoring.

Advertising