Metrohm tiamo 1.2 Manual Benutzerhandbuch

Seite 767

Advertising
background image

Subwindow Sensors

741

Order number

24 characters, [ 'empty' ]

Order number of the sensor (read-only for intelligent sensors).

Sensor serial number

10 characters, [ 'empty' ]

Serial number of the sensor (read-only for intelligent sensors).

Device

Select device, [ 'empty' ]

For non-intelligent sensors it is possible to select the device to which the
sensor is connected. All the devices in the device table which have a
measuring input are shown in the selection list. The connection of a sensor

to a particular device is not checked during the run.
For intelligent sensors the device to which the sensor is connected is
shown here automatically.

Measuring input

[ 1 ], 2

For non-intelligent sensors it is possible to select the measuring input to

which the sensor is connected. The connection of a sensor to a particular
measuring input is not checked during the run. This parameter only ap-
pears if a

Device

has been chosen.

For intelligent sensors the measuring input to which the sensor is con-

nected is shown here automatically.

Comment

24 characters

It is possible to enter remarks about the sensor.

Set to work

Select date

The current date is automatically entered here for a newly added sensor.
When editing an existing sensor the date can be selected by pressing

in the dialog window

Select date

(see General program functions - Edit - Se-

lect date).

Sensor monitoring

on, [ off ]

If this option is switched on then the working life of the sensor will be moni-
tored.

Working life

0 ... [ 999 ] days

Working life of the sensor in days. If a value is entered here then the

Expiry

date

will be adapted automatically.

Expiry date

Select date, [ Set to work + 999 days ]

Expiry date of the sensor. This date can be selected by pressing

in the

dialog window

Select date

(see General program functions - Edit - Select

date). If a date is entered here then the value for the

Working life

will be

adapted automatically.

Action

If during sensor monitoring it is found that the working life has expired then one of
the following actions will be triggered automatically at Start test:

Record message

The message that the working life of the sensor has expired will be auto-

matically saved with the determination.

Display message

A message will appear and you can choose whether to continue with the
run or cancel it. If the run is continued then the message that the working

Advertising