438

Method

Open the dialog window

Evaluation window #

(see below) to edit the pa-

rameters of the selected time window.

Delete the selected time window.

Evaluation window

Evaluation windows are defined time segments in which the dosing rate is deter-
mined by linear regression.

Lower limit

[ 0.0 ] ... 99999.9 s

Lower limit of evaluation window.

Upper limit

0.0 ... [ 999999.9 ] s

Upper limit of evaluation window.

Fix end point evaluation

on, [ off ]

For a fixed quantity (measured value, volume or time) for the fix end point
the associated values are interpolated from the list of measuring points.

Fixed quantity

[ Measured value ], Time, Volume

Selection of the fixed quantity to which the associated values for the other
quantities are to be interpolated from the list of measured points.

Fix EP# at pH

-20.000 ... 20.000, [ off ]

Fix end point 1 ... 9 for

Fixed quantity

=

Measured value

.

Fix EP# at

0.0 ... 999999.9 s, [ off ]

Fix end point 1 ... 9 for

Fixed quantity

=

Time

.

Fix EP# at

0.00000 ... 9999.99 mL, [ off ]

Fix end point 1 ... 9 for

Fixed quantity

=

Volume

.

With this button additionally to the always shown fix EP1...3 the fix EP4...9

are displayed.

With this button instead of fix EP1...9 only the fix EP1...3 are shown.

Minimum evaluation

on, [ off ]

For the minimum measured value the associated volume, time and tem-

perature are interpolated from the list of measuring points (see Titration
commands - Evaluation - Minimum/maximum evaluation
).