Method

Min. rate

0.01 ... [ 5.00 (50 µL/min)] ... [ 10.00 (100 µL/min) ] ... [ 40.00 (500 µL/min) ]
... 9999.00 µL/min

Speed of addition at the start and finish of the titration. The slower the

minimum rate that is selected, the slower the titration and the higher the ac-
curacy. the longer it takes until the control point is reached for the first time.
The following rule of thumb should be considered in order to obtain as con-
stant a dosing at the control point as possible:

Min. rate

in µL/min = Ex-

pected rate of reaction in µL/min / 10.

Note
Outside the control range the dosing rate is mainly characterized by the

Max. ra-

te

. The parameters

Max. rate

and

Dynamics

should be optimized together in such

a way that the titration does not overshoot too much when reaching the control
point. The control range should be set so that the measured value remains
within this range while it is kept constant. A slightly larger control range should
be defined for slow reactions (e. g. U = 180 mV). A SET pretitration is frequently

used to reach the control point before the substrate is added. This means that
the STAT titration can be started with a small variation from the control point.
Within the control range the dosing rate is mainly characterized by the

Min. rate

.

Titration parameters

Parameters defining the run of the titration.

Command name

25 characters, [ STAT U # ]

Name of the command.

Titration direction

Selection of the titration direction.

+

Positive measured value alteration.

[ - ]

Negative measured value alteration.

auto

The titration direction is determined automatically from the start measured
value and the set control point.

Temperature

-20.0 ... [ 25.0 ] ... 150.0 °C

(Titrando)

-170.0 ... [ 25.0 ] ... 500.0 °C

(Titrino)

Manually entered titration temperature. If a temperature sensor is con-
nected and the

Temperature measurement

in the tab

General/Hardware

under