Metrohm tiamo 1.2 Manual Benutzerhandbuch

Seite 496

Advertising
background image

470

Method

I(pol)

-125.0 ... [ 5.0 ] ... 125.0 µA

(Titrando, increment = 2.5

µ

A)

-127 ... [ 5 ] ... 127 µA

(Titrino, increment = 1

µ

A)

The polarization current is the current applied to the polarized electrode
during a voltametric measurement.

Electrode test

on, [ off ]

Switch on/off the electrode test for polarized electrodes. This test is carried
out during the transition from an inactive condition to a measuring condi-

tion. It is checked that the electrode is properly connected and that no
short-circuit is present.

Temperature measurement

Type of temperature measurement (Titrando only):

continuous

A temperature sensor must be connected. The temperature is then meas-

ured continuously.

[ automatic ]

If a temperature sensor is connected, the temperature will be measured
continuously. Otherwise the temperature manually entered under

Tempera-

ture

(see Measuring parameters) will be used.

off

The temperature will not be measured. The temperature manually entered
under

Temperature

(see Measuring parameters) will be used.

Stirrer

Stirrer

[ 1 ] ... 4, off

(Titrando only)

Selection of the stirrer.

off

means that no stirrer will be used.

Stirring rate

-15 ... [ 8 ] ... 15

(Titrando only)

Setting the stirrer speed. The direction in which stirring takes place changes
as the sign in front of the stirrer speed changes.

Switch off automatically

[ on ], off

(Titrando only)

The stirrer is automatically switched off at the end of the command if this
option is activated.

Measuring parameters

Screenshot and parameters: see MEAS U - Measuring parameters

Evaluations

Screenshot and parameters: see MEAS U - Evaluations

Additional measured values

Screenshot and parameters: see DET pH - Additional measured values

Advertising