684

Configuration

Next GLP test

Date selection, [GLP test date + 999 days ]

Date on which the next GLP test is to be carried out. This date can be se-
lected after pressing

in the dialog window

Select date

(see General

program functions - Edit - Select date). After the date has been entered the
value for the

GLP test interval

will be adapted automatically.

Action

If the time interval for the GLP test has expired then one of the following actions
will be triggered at both the Start test and the command sequence:

Record message

The message that the GLP test interval has expired will be saved automati-
cally in the determination.

Display message

A message appears and you can then choose whether to continue with the
run or to terminate it. If the run is continued then the message that the GLP

test interval has expired will be saved automatically in the determination.

Stop determination

The running determination will be automatically terminated. The following
message must be confirmed with

[OK]

.



778/789 Sample Processor

The parameters for the 778 Sample Processor and 789 Robotic Sample
Processor XL
are set on the following tabs:

General

General device information such as device name, device type, serial num-
ber, etc.

Tower #

Properties of Tower 1 and Tower 2 (if present).

Rack

Information about the attached rack.

MSB #

Properties of MSB connections 1...3 and the peripheral devices connected
to them.

RS232

Selection of the serial interface to which the device is connected.

GLP

Information about GLP test and GLP monitoring.