686

Configuration

Opens the dialog window

Load new program version

.

Old version

Shows the old device program version.

New version

Shows the new device program version to be loaded.

Loads new program version.

Note

Make sure that the device is not manipulated or switched off during the loading
process and follow the instructions shown.

Tower #

Tower parameters

Max. stroke path

0 ... [ 235 ] mm

Entry of the lowest permitted lift position. A lift height of

0 mm

corresponds

to the "Rest position", i.e. the lift is moved right to the top.

Min. beaker radius

1.0 ... 100.0 mm, [ off ]

Definition of the minimum radius which the beakers used on the rack must

have. If the lift is moved to the work position the beaker radii defined in
Rack table for general Sample positions and Special beaker will be com-
pared with the

Min. beaker radius

. If this minimum beaker radius is undercut

then the sequence will be stopped and an error message will appear. With

off

no check will be made.

Lift rate

3 ... [ 25 ] mm/s

Lift speed for manual operation.