Method

Extraction time

0 ... 999999 s

(only Titrando and 719, 720, 736, 751, 758, 785, 794, 798,

799)
The titration is not stopped until the extraction time has elapsed (even when
the EP has already been reached). For example, the entry of an extraction
time is a good idea for the titration of sparingly soluble samples.

Temperature

-20.0 ... [ 25.0 ] ... 150.0 °C

(Titrando)

-170.0 ... [ 25.0 ] ... 500.0 °C

(Titrino)

Manually entered titration temperature. If a temperature sensor is con-
nected and the

Temperature measurement

in the tab

General/Hardware

under

Sensor

is set to

automatic

or

continuous

then the temperature will be meas-

ured continuously.

Time interval measuring point

0.1 ... [ 2.0 ] ... 999999 s

(Titrando)

0.08 ... [ 2.00 ] ... 16200 s

(Titrino)

Time interval for entering a measuring point in the list of measuring points.

Stop conditions

Conditions for stopping the titration. The stop conditions are checked at the same

time and the condition which is fulfilled first stops the titration.

Command name

25 characters, [ SET pH # ]

Name of the command.

Stop volume

0.00000 ... [ 100.000 ] ... 9999.99 mL, off

(Titrando)

0.00 ... [ 100.00 ] ... 9999.99 mL, off

(Titrino)

Stops when the given volume (including start conditions) has been added

after the start of the titration. The stop volume should be adapted to suit the
sample weight or the titration vessel size in order to prevent it from over-
flowing.

Stop time

0 ... 999999 s, [ off ]

(only Titrando)

Stops when the preset time (including start conditions) has elapsed since

the start of the titration.

Filling rate

0.01 ... 166.00 mL/min, [ maximum ]

(Titrando)

0.01 ... 150.00 mL/min, [ maximum ]

(Titrino)

Speed with which the buret is to be refilled after the titration. The maximum

filling rate depends on the cylinder volume of the exchange unit or dosing
unit used.