Commands

475

Measuring time

0 ... [ 120 ] ... 999'999 s

(Titrando)

0 ... [ 120 ] ... 9'999 s

(Titrino)

Maximum period of time for measurement.

Measuring interval

0.1 ... [ 2.0 ] ... 999'999.0 s

(Titrando, increment = 0.1 s)

0.08 ... [ 2.0 ] ... 16'200 s

(Titrino, increment = 0.08 s)

Time interval for entering a measuring point in the list of measuring points.

Stop measured value

-200.0 ... 200.0 µA, [ off ]

(Titrando only)

Stop when the preset measured value has been reached since the start of
the measurement.

off

means no stop.

Temperature

Temperature

-20.0 ... [ 25.0 ] ... 150 °C

(Titrando)

-170.0 ... [ 25.0 ] ... 500 °C

(Titrino)

Manually entered measuring temperature. If a temperature sensor is con-
nected and

temperature measurement

on the tab General/Hardware is set to

automatic

or

continuous

, the temperature is measured continuously. The

value is used for the temperature adjustment during the measurement.

Evaluations

The following methods for evaluating measuring curves are defined on this tab.

Command name

25 characters, [ MEAS Upol # ]

Name of the command.

Fix end point evaluation

on, [ off ]

To a fixed quantity (measured value or time) the associated values for the

other quantities are interpolated from the list of measuring points for the
fixed end point.