Metrohm tiamo 1.2 Manual Benutzerhandbuch

Seite 453

Advertising
background image

Commands

427

Filling rate with different cylinders

When you use two dosing devices with different cylinder volumes (Dosing

cylinder 2 > Dosing cylinder 1) the filling rate of the larger cylinder must be
at the minimum:

Filling rate 2 >= Filling rate 1 • (V

Dosing cylinder 2

/ V

Dosing cylinder 1

)

Rule for dosing rate

The dosing rate must not exceed 75 % of the filling rate of the smaller cylin-
der. This corresponds to the following values (at the maximum filling rate):

Cylinder volume

Max. dosing rate

Max. flow rate

Exchange unit

Dosing unit

Exchange unit

Dosing unit

1 mL

2.25 mL/min

---

ca. 130 mL/h

---

2 mL

---

5.00 mL/min

---

ca. 300 mL/h

5 mL

11.25 mL/min

12.50 mL/min

ca. 670 mL/h

ca. 750 mL/h

10 mL

22.50 mL/min

25.00 mL/min

ca. 1.3 L/h

ca. 1.5 L/h

20 mL

45.00 mL/min

50.00 mL/min

ca. 2.7 L/h

ca. 3.0 L/h

50 mL

112.50 mL/min

124.50 mL/min

ca. 6.7 L/h

ca. 7.5 L/h

Sensor

Measuring input

[ 1 ] ... 2

(835, 836, 842, 857)

[ 1 ]

(855)

[ 1 ] ... 2, diff.

(Titrino)

Selection of the measuring input to which the sensor is connected.

Sensor

Sensor name, [ pH electrode ]

Selection of a sensor of type

pH electrode

from the list of sensors defined in

the sensor table. The calibration data for the sensor will be adopted for the
determination.

Temperature measurement

Type of temperature measuring (only Titrando):

continuous

A temperature sensor must be connected. The temperature is measured
continuously.

[ automatic ]

If a temperature sensor is connected then the temperature will be meas-

ured continuously. Otherwise the manually entered

Temperature

(tab

Titra-

tion parameters

) will used.

off

The temperature will not be measured. The

Temperature

(see Titration pa-

rameters) which has been entered manually will be used.

Stirrer

Stirrer

[ 1 ] ... 4, off

(only Titrando)

Selection of the stirrer.

off

means that no stirrer will be used.

Stirring rate

-15 ... [ 8 ] ... 15

(only Titrando)

Setting the stirrer speed. The direction in which stirring takes place changes

as the sign in front of the stirrer speed changes.

Switch off automatically

[ on ], off

(only Titrando)

If this option is enabled, the stirrer will be switched off automatically when
the command is finished.