391

Fix EP# at

0.0 ... 999999.9 s, [ off ]

Fix end point 1 ... 9 for

Fixed quantity

=

Time

.

Fix EP# at

0.00000 ... 9999.99 mL, [ off ]

Fix end point 1 ... 9 for

Fixed quantity

=

Volume

.

With this button additionally to the always shown fix EP1...3 the fix EP4...9
are displayed.

With this button instead of fix EP1...9 only the fix EP1...3 are shown.

Minimum evaluation

on, [ off ]

For the minimum measured value the associated volume, time and tem-
perature are interpolated from the list of measuring points (see Titration
commands - Evaluation - Minimum/maximum evaluation
).

Threshold value

0.1 ... [ 1.0 ] ... 20.0 pH/mL

The evaluation begins as soon as the slope of the curve exceeds the set
threshold value.

Maximum evaluation

on, [ off ]

For the maximum measured value the associated volume, time and tem-

perature are interpolated from the list of measuring points (see Titration
commands - Evaluation - Minimum/maximum evaluation
).

Threshold value

0.1 ... [ 1.0 ] ... 20.0 pH/mL

The evaluation begins as soon as the slope of the curve exceeds the set
threshold value.

Additional measured values

Screenshot and parameters: see DET pH - Additional measured values