520

Method

Temperature measurement

Type of temperature measuring (only Titrando):

continuous

A temperature sensor must be connected. The temperature is measured
continuously.

[ automatic ]

If a temperature sensor is connected then the temperature will be meas-
ured continuously. Otherwise the manually entered

Temperature

(tab

Titra-

tion parameters

) will used.

off

The temperature will not be measured. The

Temperature

(see Dosing pa-

rameters) which has been entered manually will be used.

Stirrer

Stirrer

[ 1 ] ... 4, off

(only Titrando)

Selection of the stirrer.

off

means that no stirrer will be used.

Stirring rate

-15 ... [ 8 ] ... 15

(only

Titrando)

Setting the stirrer speed. The direction in which stirring takes place changes
as the sign in front of the stirrer speed changes.

Switch off automatically

[ on ], off

(only

Titrando)

The stirrer is automatically switched off at the end of the command if this
option is enabled.

Dosing parameters

The volume to be added, the dosing rate and the dosing time are defined on this
tab.

Command name

25 characters, [ DOS pH # ]

Name of the command.

Dosing criterion

[ Volume/Dosing rate ], Volume/Dosing time, Dosing rate/Dosing time

Selects the two parameters that can be preset for dosing. The third pa-

rameter is calculated automatically from the formula

Volume = dosing time *

dosing rate

and is not shownDOS - monitored dosing.

Volume

0.00000 ... [ 10.0000 ] ... 99999.9 mL

Fixed volume to be added. This parameters is only displayed for

Dosing cri-

terion =

Volume/Dosing rate

or

Volume/Dosing time.